fluidlab.instruments.multiplexer.lakeshore_224¶
lakeshore_224¶
- class fluidlab.instruments.multiplexer.lakeshore_224.Lakeshore224(interface=None)[source]¶
Driver for the Lakeshore Model 224 Temperature Monitor
- upload_curve(curve_number, curve_name, curve_sn, curve_format, limit_value, coefficient, sensor_values, temperature_values)[source]¶
Upload a calibration curve.
- Parameters:
curve_number (int) – Specifies which curve to configure. Valid entries: 21-59.
curve_name (str) – Specifies curve name. Limited to 15 characters.
curve_sn (str) – Specifies the curve serial number. Limited to 10 characters.
curve_format (
CurveFormat
) – Specifies the curve data format. Valid entries: 1 = mV/K, 2 = V/K, 3 = Ohm/K, 4 = log Ohm/K.limit_value (float) – Specifies the curve temperature limit in kelvin.
coefficient (
CurveCoefficient
) – Specifies the curve temperature coefficient. Valid entries: 1 = negative, 2 = positive.sensor_values – Specifies sensor units, to 6 digits.
temperature_values (
numpy.ndarray
) – Specifies the corresponding temperature in kelvin, to 6 digits
- curve_number Specifies the curve an input uses for temperature conversion = <IntValue object>¶
- sensor_value Reads sensor signal in sensor units = <FloatValue object>¶
- temperature() -> Reads temperature (in Kelvins) = <FloatValue object>¶
Functions
|
Classes
|
Driver for the Lakeshore Model 224 Temperature Monitor |